Section 3 Technical Specifications
TYPICAL TECHNICAL DATA
Enclosure: Fiberglass Reinforced Polyester
Dimensions (HxWxD): 750mm x 500mm x 320mm
Enclosure height may increase to 1000mm, depending on system optional
features.
Weight: Typically 46 kg
Power Consumption: 100 W (VA)
Mains Connection: 115V AC, 60Hz or 230V AC, 50Hz (10%)
Other power options are available on request.
Mains Wire Specification: Number of Cores = 3, Current Rating minimum = 10 Amps, CSA (Cross
Sectional Area minimum) = 1.50mm
2
Signal Wire Specification: Number of Cores = 6 (+2 cores per additional signal), Current Rating
minimum = 1 Amp, CSA (Cross Sectional Area minimum) = 0.22mm
2
FEATURES IN DETAIL
Display: High Contrast 40 Character x 16 Line Backlit LCD with LED Backlight
Data Storage: Previous 9999 analysis data on screen in the microcontroller memory and
storage of data archive for the lifetime of the analyzer in the SD/MMC card
Previous 99 fault data on screen in the microcontroller memory and storage
of fault data archive for the lifetime of the analyzer in the SD/MMC card
SD/MMC Card: Flash memory card for data transfer and for software & configuration updates
Operation: Microcontroller with BioTector OS3 Software and Membrane Keyboard
Language Options: English, French, German
Other language options are available on request.
INPUT & OUTPUT SIGNALS
Standard Output: 4 programmable 4-20mA analog output signal channel fitted as standard
Maximum impedance: 500 ohms
Digital Output: 6 freely programmable system relays (volt free changeover contact with a
current rating of 1Amp at 30V DC)
One of the relays is preprogrammed as system fault relay as standard.
Data Transfer Port: SD/MMC Card
OPTIONAL FEATURES
Result Output: TIC, TC, VOC, after correlation COD, BOD
Remote Control: Input for remote start / standby
Input for remote stream and range selection (up to 2 online streams)
Input for remote manual grab sample analysis
Network Control Unit for remote access over Internet or Intranet connection
using HTTP over TCP/IP protocol
Industrial Interface: Modbus, Profibus, Ethernet
(when any of the Modbus, Profibus or Ethernet option is selected, the digital
output signals are sent through the relevant device with its specific
communication protocol)
Calibration: Automatic Calibration is available.
Calibration is not necessary between every 6 month service intervals.
Multi-stream: Up to 2 sample input channels, which can be used for online streams
Manual stream: Manual grab sample channel, which can be used for automatic calibration.
Sample input channels may be configured as manual grab sample channels.
4-20mA Outputs: As individual signal up to maximum of 4 or as multiplex signal up to maximum
of 35. Maximum impedance: 500 ohms.
Hazardous Area: Certification options are available to European Standards (ATEX for Zone 2)
and to North American Standards (Class I Division 1 and Class I Division 2).
Other options are available on request.