EasyManua.ls Logo

Hach BioTector B3500s - Installation; Basic System Requirements

Hach BioTector B3500s
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 49
Section 5 Installation
5.1 Basic System Requirements
Power and Signal Requirements
• Mains Connection: 115V AC, 60Hz or 230V AC, 50Hz (10%)
• Mains Wire Specification: Number of Cores = 3
Current Rating minimum = 10 Amps
CSA (Cross Sectional Area minimum) = 1.50mm
2
• Signal Wire Specification: Number of Cores = 6 (+2 cores per additional signal)
Current Rating minimum = 1 Amp
CSA (Cross Sectional Area minimum) = 0.22mm
2
• Power Consumption: Maximum 100 W (VA)
• Electrical Connections: Typically 4 cable glands, PG13.5, clamping range 6 - 12 mm
Air Supply and Reagent Requirements
Instrument Air Requirements
• Air Quality: - 20°C dew point (free of water, oil and dust)
To meet or exceed the air quality specification, filter pack may be
required.
• Air Supply Pressure: 1.5 bar
• Air Supply Flow Rate: Minimum 8.4 m
3
/hour at 1.5 bar
• Air Consumption: Average consumption is less than 5.4 m
3
/hour, typically 3.6 m
3
/hour
Typical Reagent Requirements
• Acid Reagent: 6.0 N Sulfuric Acid (H
2
SO
4
), containing 350 mg/l Manganese Sulphate Monohydrate.
• Base Reagent: 4.0 N Sodium Hydroxide (NaOH).
Sample, Drain and Exhaust Requirements
• Sample Inlet & Outlet Pressure: Ambient
• Sample Inlet Temperature: 2°C – 60°C (36°F - 140°F)
• Sample Flow Rate: Minimum 100ml per sample
• Sample Particle Size: Up to 100 microns
• Drain & Exhaust: Ambient

Table of Contents