Installation
1. If .NET 4.7.2 is required, unzip VAP235_NET_and_driver.zip and run NDP472-
KB405430-x86-x64-AllOS-ENU.exe.
2. If Excel 2010 or newer is not installed, install Excel 2010 or newer.
3. Run VAP235_Vx.x-BLx.xx-UIx.xx-Pxx-Pxx.exe and follow instructions.
Files and templates
After Installation Finalinspection.csv and NominalValues.csv have been installed in the
following directory: C:\ProgramData\VAP235\Files (hidden folder). This directory also
contains templates for labels and certificates. The newest NominalValue.csv file is stored
at https://danaher.sharepoint.com/:f:/r/sites/hch/team/rd/Shared%20Documents/Released
%20Software/Service%20Tools/DR300?csf=1.
Note: Update NominalValues.csv when you received a new or recalibrated test kit.
Firmware
After Installation DR300_BLx.xx_CRCxxxx.bin and DR300_UIx.xx_CRCxxxx.bin have
been installed in the application directory in the folder Update (for example: C:\Program
Files (x86)\VAP235\Update). If you get an updated file, close the program if it is opened
and replace the file right here (remove the old one).
Settings
1. Start VAP235.
2. Select the folder Settings.
3. Select Language and select English, Chinese or German.
4. Select Program settings.
5. Select the folder General. Refer to Figure 7.
6. Enter the "fi result path" and the "Certificate path" where the certificate is stored. If the
paths are not defined during installation, an error message is shown at the first start.
7. Make sure that Enable updates is activated. If enabled, VAP235 checks for new
application and boot updates that are available and necessary.
8. Push Save to save the settings.
9. Select the folder Mode and select Service.
10. At the next start, the software tool VAP235 will start with the selected settings and the
correct mode.
Inspection
17

Other manuals for Hach DR300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR300 and is the answer not in the manual?

Hach DR300 Specifications

General IconGeneral
TypePortable Spectrophotometer
Operating Temperature0 to 50 °C
MethodPre-programmed methods
Light SourceLED
Wavelength Accuracy±2 nm
Resolution0.001 Abs
DisplayLCD
Measurement MethodAbsorbance
ParameterVarious
Photometric Range0 to 2.5 A
Dimensions230 x 93 x 280 mm (9.1 x 3.7 x 11.0 in)
Photometric Accuracy±0.005 Abs