Page #33 background image
Step Description Picture
10
Insert the NG11 filter into the cell holder. Make
sure to install the filter in the correct orientation.
Install the instrument cup.
Note: The filter is light-tight. Therefore, it is not
necessary to install the instrument cap.
Push to measure the absorbance.
Make a note of the measured value of the
NG11 filter and enter the actual value in the
Certificate of Service Inspection.
11
Insert the NG5 filter into the cell holder. Make
sure to install the filter in the correct orientation.
Install the instrument cup.
Note: The filter is light-tight. Therefore, it is not
necessary to install the instrument cap.
Push to measure the absorbance.
Make a note of the measured value of the
NG5 filter and enter the actual value in the
Certificate of Service Inspection.
Inspection
31

Other manuals for Hach DR300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR300 and is the answer not in the manual?

Hach DR300 Specifications

General IconGeneral
TypePortable Spectrophotometer
Operating Temperature0 to 50 °C
MethodPre-programmed methods
Light SourceLED
Wavelength Accuracy±2 nm
Resolution0.001 Abs
DisplayLCD
Measurement MethodAbsorbance
ParameterVarious
Photometric Range0 to 2.5 A
Dimensions230 x 93 x 280 mm (9.1 x 3.7 x 11.0 in)
Photometric Accuracy±0.005 Abs