General Information
2.2 General Sensor Information
The Luminescent Dissolved Oxygen (LDO) Sensor (Figure 1) allows aqueous samples to
be easily and accurately analyzed for dissolved oxygen concentration. Specially designed
for municipal and industrial wastewater applications, the system consists of a controller
with an integrated display, and a sensor (probe with sensor cap) for in-situ measurement.
The LDO sensor can be operated using the sc100 controller and the sc1000 controller.
Refer to Operation on page 15 and sc1000 Operation on page 25 for more information.
Optional equipment, such as mounting hardware for the sensor, is supplied with an
instruction sheet for all user installation tasks. Several mounting options are available,
allowing the sensor to be adapted for use in many different applications.
Typical applications include aeration basins, nutrient removal in equalization basins,
aerobic and anaerobic digesters, effluent streams, rivers, lakes, and fish ponds.
Figure 1 LDO Sensor Dimensions
2.3 Theory of Operation
The sensor in the cap is coated with a luminescent material. Blue light from an LED
illuminates the luminescent chemical on the surface of the sensor cap. The luminescent
chemical instantly becomes excited and then as the excited chemical relaxes, it releases
red light. The red light is detected by a photodiode and the time it takes for the chemical to
return to a relaxed state is measured. The higher the oxygen concentration, the less red
light is given off by the sensor and the shorter time it takes for the luminescent material to
return to a relaxed state. The oxygen concentration is inversely proportional to the time it
takes for the luminescent material to return to a relaxed state.
Unlike electrochemical dissolved oxygen sensor technologies, the Luminescent Dissolved
Oxygen (LDO) sensor does not consume oxygen. It does not require frequent
recalibration or frequent cleaning (except when associated with consumptive slimes),
resulting in longer sensor life and more stable and accurate readings. The system is also
flow-independent so measurements can be made in applications with low or no flow.

Table of Contents

Other manuals for Hach LDO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach LDO and is the answer not in the manual?

Hach LDO Specifications

General IconGeneral
Storage Temperature-20 to 60 °C (-4 to 140 °F)
IP RatingIP68
Range0 to 20 mg/L
Accuracy±0.1 mg/L or ±1% of reading, whichever is greater
Resolution0.01 mg/L
Measurement Range0 to 20 mg/L
DescriptionLuminescent Dissolved Oxygen (LDO) sensor
Response timet95 < 30 seconds
Battery LifeN/A (powered externally)