30
Troubleshooting
Table 7 Warning Codes
Displayed
Warning
Definition Resolution
EE SETUP ERR
The EEPROM was corrupted. Values have
been set to factory defaults.
Contact the Service Department.
EE RSRVD ERR
The EEPROM was corrupted. Values have
been set to factory defaults.
Contact the Service Department.
TEMP < 0 C The sensed temperature is below 0 °C (32 °F).
Increase process temperature or discontinue use until
the process temperature is above 0 °C (32 °F).
TEMP > 50 C
The sensed temperature is above 50°C
(120 °F).
Decrease process temperature or discontinue use until
the process temperature is below 50 °C (120 °F).
RED AMPL LOW
Sensor cap not installed or not installed
correctly.
Sensor not operating properly.
Remove and reinstall sensor cap.
Contact the Service Department.
RED AMPL HIGH
Sensor cap not installed or not installed
correctly.
Sensor not operating properly.
Remove and reinstall sensor cap.
Contact the Service Department.
BLUE AMPL LOW
Sensor cap not installed or not installed
correctly.
Sensor not operating properly.
Remove and reinstall sensor cap.
Contact the Service Department.
BLUE AMPL HIGH
Sensor cap not installed or not installed
correctly. Sensor not operating properly.
Remove and reinstall sensor cap.
Contact the Service Department.