Section 5 Operation
5.1 Sensor Setup
When a sensor is initially installed, the serial number of the Digital Gateway will be
displayed as the sensor name. To change the sensor name refer to the following
instructions:
1. Select Main Menu.
2. From the Main Menu, select SENSOR SETUP and confirm.
3. Highlight the appropriate sensor if more than one sensor is attached and confirm.
4. Select CONFIGURE and confirm.
5. Select EDIT NAME and edit the name. Confirm or cancel to return to the Sensor
Setup menu.
5.2 Sensor Data Logging
The sc1000 controller provides a data log for each sensor.
The sc100 provides three data logs (one for each sensor and one for calculated values)
and three event logs (one for each sensor and one for the controller). The data logs store
the measurement data at selected intervals. The event log stores a variety of events that
occur on the devices such as alarms and warning conditions. The data logs are stored in a
packed binary format and the event logs are stored in a CSV format. The logs can be
downloaded through the digital network port, service port, or the IrDA port. DataCom is
needed for downloading logs to a computer.
1. From the Main Menu, select SENSOR SETUP and confirm.
2. Highlight the appropriate sensor if more than one sensor is attached and confirm.
3. Select CONFIGURE and confirm.
4. Select LOG SETUP and cofirm.
5. Select from the logging interval options from the list box and confirm.
5.3 Sensor Diagnostics Menu
SELECT SENSOR
ERROR LIST—See section 7.1 on page 29.
WARNING LIST—See section 7.2 on page 29.