Figure 1 Fill the antifoam cartridge
Note: After installation of the antifoam, it is critical that any air bubbles be cleared from the antifoam circuit
before taking any measurements. To do this, select the Antifoam system option from the Actuator
verification option in the Maintenance menu and follow the instructions as described in Antifoam system
on page 22. Select 200 injections and stop the process when the system constantly spits a small stream of
antifoam at each stroke denoting the antifoam circuit is free of air bubbles.
Note: A cartridge completely filled with antifoam is sufficient for approximately 3,000 measurements.
Digital input verification
This option allows you to check all devices that provide digital inputs to the instrument
microprocessor. A black dot in the circle to the right of the option signifies that the sensor is
activated.