Figure 6 Probe storage
Section 9 Troubleshooting
Keep the probe clean and in the recommended storage solution when not in use for the best
accuracy, stabilization time and life of the probe.
Problem Possible cause Solution
Decreased probe performance
causes slow stabilization and
prevents accurate calibrations
or measurements.
The glass sensor is dirty. Clean and condition the probe.
Refer to Clean the probe
on page 10.
The glass sensor has become
dry.
Clean and condition the probe.
Refer to Maintenance
on page 10.
The calibration slope of the
probe has changed.
Increase the accepted slope limit
settings if possible, or contact
technical support.
English 13