Hach sensION+ DO6 - Two-Point Calibration

Hach sensION+ DO6
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
6. Operation
ENGLISH
6.4.2. Two-point calibration
6.4.2.1. Preparation
1. Unscrew the sensor protector.
2. Dip the sensor into an oxygen-free solution.
3. Stir the sensor gently.
6.4.2.2. Calibration
Follow the diagram.
Note:
To perform a two-point calibration, you must be certain
that the solution is oxygen-free; otherwise, one-point
calibration (100% DO) using the calibrator tube is
preferable.
If calibration is not performed correctly, the membrane
must be cleaned and the electrolyte changed (see sen-
sor manual). The calibration is correct if an oxygen
content of over 98% is displayed when taking a mea-
surement in the calibrator tube. A value of over 99%
is optimal.
If the measuring instrument is not used for 5 minutes, it
shuts off automatically.
Dip the sen-
sor into an
oxygen-free
solution