Figure 8 Probe connections
Complete probe measurements
Do the procedure that follows to complete a measurement with probe(s). Refer to Figure 9. Multiple
Chemkey measurements can be completed at the same time. Refer to Complete Chemkey
measurements on page 11.
1. The probe must have the correct service-life time stamp. Set the date and time in the meter
before the probe is attached. Refer to Change the date and time on page 10.
2. If complete traceability is necessary, enter a site ID and operator ID before measurement. Refer
to Use an operator ID on page 10 and Use a site ID on page 10.
3. Remove the probe connector port caps and attach the probe(s). Refer to Figure 8 on page 13 for
the probe connection.
Note: Do not discard the probe connector port caps. Install the probe connector port caps again when no probe
is connected.
4. Rinse the sample flasks with sample.
5. Fill the sample flask with sample so that the sensor is fully in the sample.
6. Put the probe(s) into the probe sample flasks and push Read if the measurement mode is set to
Press to Read. Refer to the probe documentation for more information.
7. The measurement values are shown on the display.