Hach SL1000 - Probe verification; Chemkey verification; Data log

Hach SL1000
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Complete the instructions shown on the display to calibrate the Chemkey.
6. If necessary, do a verification. Refer to Chemkey verification on page 15.
Probe verification
1. Push Main Menu>Verification to start a verification.
2. Select the applicable device.
3. If necessary enter the applicable standard adjust value.
4. Complete the instructions shown on the display to do the verification.
Chemkey verification
There are two options to complete a Chemkey verification. Use a parameter-specific Chemkey or a
system verification Chemkey for the verification. Refer to Replacement parts and accessories
on page 21.
Parameter-specific Chemkey:
1. Push the parameter specific Chemkey slowly, in one movement, into the slot(s) 1 to 4.
2. Then, push Main Menu>Verification to start the verification.
3. Complete the menu guided process.
System verification Chemkey:
1. Push Main Menu>Verification to start a verification.
2. Then, push the system verification Chemkey slowly, in one movement, into the first slot.
3. Complete the menu guided process.
Data log
The data log shows all saved measurements.
1. Push Main Menu>Data Log.
2. Select an option.
Option Description
All Logs by Date Shows sample measurement results: The probe name, probe serial number, value,
unit, temperature, time, date, notes, operator ID, site ID and calibration details.
Current Cal on Probe Shows the calibration details for the most recent calibration. If the probe has not
been calibrated by the user, the factory calibration data is shown.
Cal History on Probe Shows a list of the times when the probe was calibrated. Select a date and time to
view a summary of the calibration data.
Delete Data Log Erases all data of the meter at once. The device data will be kept.
Import data to Excel
W A R N I N G
Electrical shock hazard. Externally connected equipment must have an applicable country safety
standard assessment.
English 15