DOC022.97.80457
SL1000
08/2014, Edition 1
Basic User Manual

Other manuals for Hach SL1000

Questions and Answers

  • C
    cruztyroneAug 19, 2025
    What to do if Hach SL1000 shows Overrange?
    • D
      Douglas CarlsonAug 19, 2025
      If the Hach Measuring Instruments displays 'Overrange', it means the concentration exceeds the method's upper limit. To resolve this, dilute the sample and measure it again.
  • D
    Dr. Tony JacksonAug 21, 2025
    Why does my Hach Measuring Instruments say 'Barcode read error'?
    • M
      murphyemilyAug 21, 2025
      If the Hach Measuring Instruments displays 'Barcode read error', it indicates that the Chemkey isn't operating or installed correctly. Try reinserting the Chemkey, using a new one, or cleaning the barcode window.
  • B
    Brandon HaleAug 23, 2025
    What to do if Hach SL1000 displays Sample aspirate error?
    • C
      Christopher BurtonAug 23, 2025
      If the Hach Measuring Instruments displays 'Sample aspirate error', it means the Chemkey isn't working correctly or isn't fully inserted. Try a new Chemkey, ensuring it's fully pushed into the slot.
  • A
    Adam JohnsonAug 26, 2025
    Why does Hach SL1000 Measuring Instruments say 'Used Chemkey'?
    • L
      lorrainesmithAug 26, 2025
      If the Hach Measuring Instruments displays 'Used Chemkey' error, it means the Chemkey has already been used, or there is water in the slot. Use a new Chemkey. If the issue continues, dry the slot and instrument before trying again.
  • M
    Marissa CoxAug 27, 2025
    How to fix 'Meter does not detect sample' on Hach SL1000 Measuring Instruments?
    • M
      Michael GoodwinAug 27, 2025
      If the Hach Measuring Instruments displays 'Meter does not detect sample', ensure the sample cup is filled to the fill line and that the sample conductivity is more than 2 µS/cm.
  • C
    Christian AdamsAug 29, 2025
    How to fix 'Low Light. Clean Chemkey slot' error on Hach SL1000?
    • S
      Samantha BaldwinAug 30, 2025
      If the Hach Measuring Instruments displays 'Low Light. Clean Chemkey slot', it indicates that the Chemkey isn't operating or installed correctly. Try reinserting the Chemkey in one push, using a new Chemkey, or cleaning the Chemkey slot.
  • K
    Kelli RivasSep 1, 2025
    What to do if Hach Measuring Instruments says 'Insert Chemkey in darker environment'?
    • D
      Dana YuSep 1, 2025
      If the Hach Measuring Instruments displays 'Insert Chemkey in darker environment. Too much ambient light (1)', it means the ambient light is too high to read the barcode. Move the meter to a darker environment and reinsert the Chemkey into the slot.
  • K
    Katie RiveraSep 3, 2025
    Why does Hach SL1000 Measuring Instruments say 'Too much ambient light'?
    • P
      Paula SanchezSep 3, 2025
      If the Hach Measuring Instruments displays 'Insert Chemkey in darker environment. Too much ambient light (2)', it means the ambient light is too high to make a valid measurement. Move the meter to a darker environment and measure again.
  • P
    Paige GeorgeSep 5, 2025
    What to do if Hach SL1000 Measuring Instruments displays Tray Leak?
    • R
      Roberta DunnSep 5, 2025
      If the Hach Measuring Instruments displays 'Tray Leak', it indicates a problem with the tray. Replace the tray.
  • J
    Jeffrey MilesSep 7, 2025
    Why does Hach SL1000 Measuring Instruments say 'Tray too cold'?
    • R
      Rachel CainSep 7, 2025
      If the Hach Measuring Instruments displays 'Tray too cold', it means the meter may be too cold to heat the tray, or the heater isn't working. Move the meter to a warmer area and measure again, or replace the tray.

Summary

Safety information

Use of hazard information

Explains the meaning of DANGER, WARNING, CAUTION, and NOTICE symbols.

Precautionary labels

Describes the precautionary labels and their referenced symbols.

Product overview

Installation

Lithium battery safety

Details critical safety precautions for handling lithium batteries.

User interface and navigation

Display description

Details the various elements and indicators shown on the measurement screen.

Keypad description

Startup

Standard operation

Complete Chemkey measurements

Complete probe measurements

Calibration

Probe verification

Chemkey verification

Data log

Import data to Excel

Maintenance

Troubleshooting

Clean or replace parts

Do a diagnostic check

Replacement parts and accessories

Hach SL1000 Specifications

General IconGeneral
Measurement RangeVaries by parameter
ResolutionVaries by parameter
AccuracyVaries by parameter
Light SourceLED
DetectorPhotodiode
DisplayLCD
ConnectivityUSB, Bluetooth
Battery Life8 hours
Operating Temperature0 to 50 °C
Measurement ParametersChlorine
MethodColorimetric
Power SupplyRechargeable battery