P/N 53063-50, Manual Revision: B Sequestering on the Cell Saver 5+ System
7-19
8. The white buffy coat band (made up of platelets and white cells), which is
immediately adjacent to the top of the red cell level, will be seen coming up
over the shoulder of the bowl. The machine bowl optics determines the
optimum time for collection of PRP. When this is determined the pump
pauses and the operator is prompted to open the BLUE (PRP) bag clamp and
close the YELLOW (PPP) bag clamp. Once this is completed the machine
waits for the START key to be pressed to continue processing as shown in
the following screen display:
Caution: Never close the yellow (PPP), blue (PRP), and white (air) bag clamps at
the same time.
125
ML
PAUSE
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to resume collection of PRP
SEQUESTER
239 ml
0 ml
0
Pump Rate: 60
Centrifuge: 4750
Press MODIFY to change parameters
ml/min
rpm
PPP Volume:
PRP Volume:
100 ml
0 ml
100/
0/
OPEN BLUE bag clamp
CLOSE YELLOW bag clamp