P/N 53063-50, Manual Revision: B Introduction
1-15
Bowl symbol
chart
During a procedure, the bowl icon on screen reflects the bowl size currently in
use. See below for bowl sizes and corresponding icons.
Primary Power a. Electrical Voltage: 110/220 VAC (±15%), switch-selectable.
b. Operating Frequency: 47 – 63 Hz.
c. Operating Current:
! At 110 Vac 2.6 amp, steady state; 4 amp, surge,
! At 220 Vac 1.3 amp, steady state; 2 amp, surge.
d. Power Cord: Detachable 16-foot (4.9m) E-183553/C, #18 low-leakage SJT P-
350, <HAR>, VDE tested.
e. The Cell Saver 5+ has UL approval.
f. Input Fuse Rating: M2.5 Amperes @ 250 VAC
Environmental
Conditions
a. Operating Temperature: 10° to 27° Celsius (50° to 80° F).
b. Humidity: 95% maximum, noncondensing.
c. Possible Explosion Hazard if used in the presence of flammable anesthetics
Storage and
Transport
Conditions
a. Relative humidity: 8% – 80% rh
b. Temperature: –20°C to + 50°C
70 mL bowl
125 mL bowl
225 mL bowl
70
125
ML
225
ML