Appendices P/N 53063-50, Manual Revision: B
APP-12
IEC 60601-1-2: 2001, Table 202: Guidance and manufacturer’s declaration - electromagnetic immunity
The Cell Saver system is intended for use in the electromagnetic environment specified below. The customer
or the user of the Cell Saver system should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient burst
IIEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
EC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips,
short
interruptions,
voltage
variations on
power supply
input lines.
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0,5 cycles
40% U
T
(60% dip in U
T)
for 5 cycles
70% U
T
(30% dip in U
T)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0,5 cycles
40% U
T
(60% dip in U
T)
for 5 cycles
70% U
T
(30% dip in U
T)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the operator of the Cell Saver system
requires continued operation during power
mains interruptions, it is recommended that
the Cell Saver system be powered from an
uninterruptible power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment
NOTE: U
T
is the AC mains voltage prior to application of the test level.