P/N 53063-50, Manual Revision: B Equipment and Disposable Description
2-7
WASH Cell Saver: Pumps saline into spinning bowl from saline bag.
Sequester: Not used.
EMPTY Cell Saver: Pumps fluid from stationary bowl into product bag or reser-
voir.
Sequester: Pumps fluid from stationary bowl into product bag or reser-
voir.
RETURN Cell Saver: Pumps fluid from stationary bowl back into reservoir or into
extracorporeal circuit. [Alternative to Empty where contents of bowl
are pumped back (returned) through “Fill” line (red) instead of (emp-
tied) to product bag.]
Sequester: Not used.
The following keys control the pump:
Cell Saver: Raises the pump speed in increments of 25 ml/min to a
maximum of 1000 ml/min (maximum of 600 ml/min in Automatic).
Sequester: Raises the pump speed in increments of 10 ml/min to a
maximum of 250 ml/min in FILL and CONC and in increments of 25
ml/min to a maximum of 1000 ml/min for EMPTY.
Cell Saver: Lowers the pump speed in increments of 25 ml/min to a
minimum of 25 (minimum of 200 ml/min in Automatic).
Sequester: Lowers the pump speed in increments of 10 ml/min to a
minimum of 10 ml/min in FILL and CONC and in increments of 25 ml/
min to a minimum of 25 ml/min in EMPTY.
PAUSE Pause: Stops the pump. Does not stop the centrifuge.
The following keys are “modify action” keys:
MODIFY Allows operator to access the modifiable parameters.
/Yes Either increases a numerical value, answers “yes” to a yes-no question,
or toggles between multiple options.
/No Either decreases a numerical value, answers “no” to a yes-no question,
or toggles between multiple options.
SELECT Moves down a list of parameters/questions.