P/N 53063-50, Manual Revision: B Automatic Operation: Cell Salvage
5-9
When air is sensed in the WASH mode and less than 90% of the wash volume
has been processed, the pump stops, an audible chime is sounded and the
display reads:
The operator should replace the saline bags and then press START to resume
processing.
If 90% or more of the wash volume has been pumped, then the system advances
to EMPTY.
Warning: At anytime during the Automatic operation the operator can advance
the system to MANUAL operation by pressing the MODE key.
Because of the controls and “intelligence” built into the machine, automatic
processing will produce consistently good results
Warning: Results are dependent upon the correct setup of solutions and
disposables.
Saline bag empty
Replace saline bag