Notice Messages and Troubleshooting 5-23
P/N 39551-00, Rev. F
440
SELF TEST PROBLEM
LOW PRESSURE ERROR AT THE DPM.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
442
SELF TEST PROBLEM
HIGH PRESSURE ERROR AT THE DPM.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
443
SELF TEST PROBLEM
STATE TEST ON CHANNEL 1 FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
444
SELF TEST PROBLEM
STATE TEST ON CHANNEL 2 FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
447
SELF TEST PROBLEM
TIME BASE TEST ON CHANNEL 1 FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
448
SELF TEST PROBLEM
TIME BASE TEST ON CHANNEL 2 FAILED.
1. Turn power OFF, then ON and try again
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
449
452
453
SELF TEST PROBLEM
THE SAFETY SYSTEM FAILED THE SELF TEST.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
454
VALVE PROBLEM
PURPLE VALVE MAY NOT BE LOADED.
The disposable test indicates
that the valve may not be loaded.
1. Check tubing installation at valve.
2. Resume operation.
455
VALVE PROBLEM
ORANGE VALVE MAY NOT BE LOADED.
The disposable test indicates
that the valve may not be loaded.
1. Check tubing installation at valve.
2. Resume operation.
456
VALVE PROBLEM
RED VALVE MAY NOT BE LOADED.
The disposable test indicates
that the valve may not be loaded.
1. Check tubing installation at valve.
2. Resume operation.
458
VALVE PROBLEM
WHITE VALVE MAY NOT BE LOADED.
The disposable test indicates
that the valve may not be loaded.
1. Check tubing installation at valve.
2. Resume operation.
459
VALVE PROBLEM
YELLOW VALVE MAY NOT BE LOADED.
The disposable test indicates
that the valve may not be loaded.
1. Check tubing installation at valve.
2. Resume operation.