5-24 Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
460
VALVE PROBLEM
GREEN VALVE MAY NOT BE LOADED.
The disposable test indicates
that the valve may not be loaded.
1. Check tubing installation at valve.
2. Resume operation.
461
SELF TEST PROBLEM
THE CENTRIFUGE VACUUM SYSTEM IS OBSTRUCTED.
Centrifuge vacuum could not be released.
1. Remove bowl.
2. Check that white filter in Centrifuge
Chuck base is dry.
3. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
462
SELF TEST PROBLEM
DISPOSABLE ID READER FAILED SELF TEST.
1. Remove white pump manifold.
2. Clean disposable ID reader window.
3. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
466
WEIGHER PROBLEM
PLEASE REMOVE ALL WEIGHT FROM WEIGHER.
1. Remove any weight on Weigher.
2. Resume operation.
3. Replace bag AFTER self test.
467
SELF TEST PROBLEM
PLEASE DISCONNECT THE CUFF FROM REAR PANEL.
Excessive Cuff pressure was measured.
1. Disconnect cuff from rear panel.
2. Deflate cuff completely.
3. Connect cuff to rear panel.
4. Resume operation.
468
SELF TEST PROBLEM
IS LINE SENSOR CLEAN AND TUBING REMOVED?
1. Remove tubing from Line Sensor.
2. Check Line Sensor for contamination.
3. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
469
SELF TEST
Please OPEN the Centrifuge Cover Latch.
The Centrifuge Cover Switch must be
tested for safe operation.
1. OPEN and CLOSE the Centrifuge Cover
Switch as requested.
470
SELF TEST
Please CLOSE the Centrifuge Cover Latch.
The Centrifuge Cover Switch must be
tested for safe operation.
1. OPEN and CLOSE the Centrifuge Cover
Switch as requested.
471
NOTICE
THE PROTOCOL PARAMETERS HAVE BEEN RESET.
The saved protocol parameters have been
reset. Before starting the next procedure:
1. Reprogram the HaemoCalculator.
2. Reprogram Modify.
SOFTWARE WATCHDOG
INVALID ID.
The computer has sensed an
invalid Software Watchdog ID code.
Please turn the power off and
contact Haemonetics Technical Hot Line.
# Notice Messages Help Messages

Table of Contents

Other manuals for Haemonetics MCS+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics MCS+ and is the answer not in the manual?

Haemonetics MCS+ Specifications

General IconGeneral
BrandHaemonetics
ModelMCS+
CategoryMedical Equipment
LanguageEnglish

Related product manuals