Use screened cables, max 3 meters, not outside buildings.
Short as possible, max 3 meters, screened, proper ground.
High frequency fields may cause deviations in readings.
RF fields can affect measurements, coupling via test leads or mains.
Vertical input channels, operating modes, sensitivities, and bandwidth.
Time base accuracy, sweep modes, time/division ranges, and X-Y phase shift.
Trigger modes, sources, coupling, slope, and active TV sync separation.
Sampling rate, storage, refresh, dot join, peak detect, and memory.
Setup, autoset, save/recall, and readout/cursor functions.
PC remote control options and PC interface details.
Test voltage, current, and frequency for component testing.
CRT details, calibration, power, temperature, weight, and dimensions.
Diagram showing board placement from the top of the oscilloscope.
Diagram showing board placement from the back of the oscilloscope.
Diagram showing board placement from the bottom of the oscilloscope.
Adjust TB 50ns 5mV/Div 1MHz squarewave 25mVPP.
TB 200ns 5mV/Div 100kHz/1MHz squarewave 25mVPP.
Set Start A 2mm before Start B.
Set Start B equal to Start A.
Adjust X/Y CH2 5mV/Div 10kHz squarewave 25mVPP.
Set X-Pos. to center, adjust beam length.
Adjust TB 1μs MAG10 10MHz sine wave 7mVeff.