10
Subject to change without notice
Measurement setup 2
This setup is used whenever a frequency source inside the
HM5014-2 is to measured to determine the exact frequency.
The frequency counter [M4] has to be connected to an external
reference with a minimum accuracy of 1E-9, if not otherwise
specifi ed.
1. Basics
1.3 Measurement Equipment and Accessories
Item Type of Equipment Equipment minimum
requirements
Recommended Equipment Purpose
M1 RF Signal Generator HM8134-3 with inter-
face
HM8135 with interface Used as signal source in multiple
procedures
M2 Power Meter Rohde&Schwarz
NRVS
Rohde&Schwarz NRVS Tracking Generator adjustment,
Test Signal adjustment
M3 Insertion Unit (for M2) 100 kHz to 1.05 GHz Rohde&Schwarz NRV-Z5 or
Rohde&Schwarz NRV-Z4
Tracking Generator adjustment,
Test Signal adjustment
M4 Frequency Counter Frequency Measure-
ment Accuracy: 0.001
ppm, External Refer-
ence
HAMEG HM8123 HM5014-2 TCXO and XO adjustment
M5 N Coaxial cable 50 Ohm, DC-8 GHz,
N connectors (male/
male) Length 1 – 1.5 m
Huber&Suhner No.
23023806
Signal interconnection
M6 SMB /BNC Coaxial Adap-
tor
50 Ohm, SMB male,
BNC female
Schuricht 33-SMB-
BNC-50-2 (Huber+Suhner)
Signal interconnection
M7 RS-232 cable Connector 9-pin Sub-D
(male/female), length
1-2.5 m
HAMEG HZ14 Interface connection, PC to
HM5014-2
M8 Coaxial Cable BNC-BNC BNC male Connec-
tor both ends, length
1-1.5 m
HAMEG HZ34 Signal interconnection
M9 Coaxial Cable
N to SMA
50 Ohm, DC … 6 GHz,
N-connector (male),
SMA connector (male),
length 1 – 1.5 m
Huber&Suhner No.
23005050
Connection of TG output to frequen-
cy counter
M10 Digital Multimeter 3 1/2 digit resolution HM8012 Measurement of probe voltage
M11 Adaptor N-BNC 50 Ohm N male, BNC
female
HZ21 Signal interconnection
Table 1-1
Measurement setup 1
This setup is used whenever a signal has to be provided for
the HM5014-2. It is used in multiple steps during checking
and adjustment. The synthesizer [M1] has to be connected to
an external reference with a minimum accuracy of 1E-9, if not
otherwise specifi ed.
RF OUTPUT
RF INPUT
N cable
(M5)
HM8134-3
(M1)
HM5014-2
Figure 1.3.1: Measurement SETUP 1
Input A
BNC cable
(M8)
Adapter
(M7)
HM5014-2
HM8123
(M4)
Figure 1.3.2: Measurement SETUP 2
Input C
A
B