26
Lift does not move in one direction
Code: Flat line (on charge) Code: 2 (off charge)
The voltage between M7-2 and ground is not 4,5 V
The voltage between M7-1 and ground is not 4,5 V
when moving the toggle
The voltage between M7-3 and ground is 4,5 V
When moving the toggle
The main board P1 is faulty
There is break in the current between:
M7-1, U1-3, U2-3, U6-3, U12-3, U13-3
There is break in the current between:
M7-3, U1-1, U2-1, U6-1, U12-1, U13-1
The joystick is faulty This Main Board P1 is faulty
The main board P1 Is faulty The joystick is faulty
An armrest switch S29 or S30 is faulty An armrest switch S29 or S30 is faulty
Connect the molex testing link