17
A “PLEASE WAIT” message will be displayed followed by the number of
free spaces. During active interval logging, lot information is displayed
on the third LCD line. The line indicates in which lot the data will be
placed and keeps track of the number of logged records taken. The
“LOG“ tag is continuously on during active logging.
Pressing RANGE/u key during an interval log will display the number
of logs available.
Pressing the LOG key again will stop the Interval logging session. The
“LOG STOPPED“ message will be displayed for a few seconds.
If a sensor failure occurs during interval logging, the message “OUT OF
SPEC.” will alternate with logging information.
Manual Logging
Select Manual in the SETUP menu. To initiate a Manual log, press the
LOG key while the instrument is in measurement mode. The “PLEASE
WAIT” screen will be displayed briefly followed by a screen indicating
the measurement has been saved and then a screen indicating the log
record number.
The “LOG” tag will be displayed on all 3 screens.
“PLEASE WAIT
“SAVED” with the log record number
“FREE” with the number of free spaces available
SETUP/INSTALLATION