8
DIAGRAM
KEYPAD FUNCTION
1. CAL/MODIFY ‑ Used to enter and exit calibration mode. In SETUP, used to initiate changes of a
configuration setting.
2. GLP/CFM ‑ Used to display GLP calibration information. In SETUP, used to confirm change made.
In calibration, used to accept calibration points.
3. RANGE/u ‑ Used to select measurement range. In SETUP, used to move to right in pick list. In
log RCL, used to view GLP data for a data point.
4. SETUP/CLR ‑ Used to enter/exit SETUP mode. During calibration, used to clear previous
calibration data. In log RCL, used to clear log records.
5. / ‑ Used to scroll through SETUP menu. Used to change selection when modifying a
parameter in SETUP.
6. RCL (Recall) ‑ Used to view logged records or view % log memory used.
7. LOG ‑ Used to log data by manual log on demand or manual log on stability or to
start/stop interval logging.
Note: You can increase/decrease the speed to change the value of a parameter.
Proceed as follows:
Press and hold down the or key, then slide the finger toward the double
apex to increase the speed that a value changes.