33
SPECIFICATIONS
Dissolved Oxygen Specifications
DO
Range
0.00 to 45.00 ppm (mg/L);
0.0 to 300.0%;
Resolution
0.01 ppm (mg/L);
0.1%;
Accuracy
@ 25 °C / 77 °F
±1.5% of reading ±1 digit;
Calibration One or two points at 0% (HI7040) and 100% (water saturated air)
Temperature
Compensation
0.0 to 50.0 ºC; (32.0 to 122 ºF)
Salinity
Compensation
0 to 40 g/L;
(with 1 g/L resolution)
Altitude
Compensation
‑500 to 4,000 m (‑1640 to 13120’);
(with 100 m (328’) resolution)
Probe
HI764080
Log Feature Up to 1000 records organized in:
Manual Log on Demand (Max. 200 logs)
Manual Log on Stability (Max. 200 logs)
Interval Logging (100 lots, Max. 600 logs/lot)
Temperature
Range 20.0 to 120.0 ºC ; ‑4.0 to 248.0 °F
Resolution 0.1 ºC; 0.1 ºF
Accuracy ±0.5 ºC; ±0.9 ºF
Additional
Specifications
PC Interface Micro USB
Storage Interface USB
Power Supply
5 VDC Adapter (included)
Environment
0‑50 ºC (32122 ºF) Max 95% RH non‑condensing
Dimensions
202 x 140 x 12 mm (7.9 x 5.5 z 0.5”)
Weight
250 g (8.82 oz)