Operational Guide
26
Wrong Standard Temperature
If the temperature is out of the 0.0 to 60.0 ºC range,
“WRONG STANDARD TEMPERATURE” message will be
displayed and the temperature value will blink.
GLP is a set of functions that allows storage and retrieval of data regarding the
maintenance and status of the electrode. The following information can be
viewed on the third LCD line during measurement:
TEMPERATURE SENSOR problem (if there is one)
CAL DUE or CELL FACTOR
T.Coef. or T.Ref. (User selected)
TIME
DATE
RANGE
BATTERY or CHARGE STATUS
To view more information, enter GLP mode by pressing GLP key.
EC calibration data is stored automatically after a successful calibration.
To view the EC calibration data, press GLP when the instrument is in EC
measurement mode. The instrument will display the calibration standard and
the temperature of the calibrated standard. Use the pq keys to scroll through
the calibration data displayed on the third LCD line.
The cell factor in cm
-1
determined from the calibration
with the current reading.
The calibration o set factor in μS/cm together with
the current reading.
The Solution Standard and calibration temperature.
The temperature coe cient used during calibration
with the current reading.
Calibration
Messages
EC/TDS GLP
Information

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments edge EC HI2003 and is the answer not in the manual?

Hanna Instruments edge EC HI2003 Specifications

General IconGeneral
BrandHanna Instruments
Modeledge EC HI2003
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals