Submerse the temperature probe in the second vessel as near as
possible to the reference thermometer. Allow a few seconds for the
probe to stabilize.
Use the ARROW keys to set the reading on the secondary LCD to
that of the hot water.
When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
Press CFM to confirm. The instrument returns
to measurement mode.
Note: If the reading is not close to the selected calibration point,
WRONG” tag will blink. Change the temperature probe and
restart calibration.