48
• Using the arrow keys, set the
controller address. To exit setup,
press CFM key and then LCD key.
Note In an RS 485 network, each device must have a different
address.
• Select the desired baud rate with the up and down keys
and press the CFM key to store the new settings. Available
baud rates are: 1200, 2400, 4800 and 9600 Bps.
The PC software to communicate with the controllers is
HI92500. Please refer to this product specifications for fur-
ther informations.
RS485 PROTOCOL
The commands sent to the controller must have the follow-
ing format:
• 2-digit process ID number
• 3-character command name
• Parameters (variable length, may be null)
• End of command (always the CR character, Hex 0D)
A maximum time interval of 20 ms is allowed between two
consecutive characters of a command. It is possible to send
commands to change the controller settings or to simply
ask information on the controller status. Following is the
complete list of commands available:
CommandCommand
CommandCommand
Command
ParameterParameter
ParameterParameter
Parameter
DescriptionDescription
DescriptionDescription
Description
CAR null Request calibration data
GET NN Request setup item NN
K01 null Same as CFM+CAL keys
K02 null Same as LCD+CAL+SETUP keys
KCD null Same as CAL DATA key
KCF null Same as CFM key
KCL null Same as CAL key