49
KDS null Same as LCD key
KDW null Same as key
KRG null Same as key
KST null Same as SETUP key
KUP null Same as key
MDR null Request firmware code
ECR null Request EC reading (available in
control or idle mode only)
TDR null Request TDS reading (available
in control or idle mode only, for
HI 24 series only)
RNG null Request measure Range
(available in control
or idle mode only)
TMR null Request temperature reading
PWD NNNN Send the 4-digit password
SET NNPC
1
C
2
C
3
C
4
C
5
Set setup item NN to the PC
1
C
2
C
3
C
4
C
5
value; P=+ if the value is greater than 0, P=- if the value
is less than 0. C
1
can be 0 or 1; only C
2
C
3
C
4
C
5
can be 0 to 9
or blank (the command is not available if the controller is in
setup mode)
Note If the controller is not in control or idle mode and the tem-
perature reading is requested through the TMR command,
the controller answers with the last acquired reading when it
was in control or idle mode.
Following are examples of commands for setup items:
“03SET22+01200<CR>”“03SET22+01200<CR>”
“03SET22+01200<CR>”“03SET22+01200<CR>”
“03SET22+01200<CR>”
This command sets the setup item 22 (relay 2 set point) of a
EC controller, identified by the process ID number 03, to the
+12.00 mS value.