15
••
••
• Take the HI 93703-0
bottle
containing the ZERO FTU*
Standard (or turbidity-free
dilution water) and fill the
measurement cuvette.
Note: In order to minimize any error introduced by the cuvette,
it is recommended to use, during calibration, the same cuvette
you are going to use to perform the measurement.
••
••
• Insert the cuvette with the
HI 93703-0 @ ZERO FTU*
standard solution (or turbidity-
free dilution water) into the
measurement cell and press the
CAL key. A blinking “SIP“
message indicates that the
instrument is performing the
measurement. After approximately
50 seconds the instrument will
ask for the HI 93703-10
standard solution @10 FTU*
by displaying “10.0“.
••
••
• Repeat the same procedure with
HI 93703-10 @10 FTU*
standard solution.
••
••
• Insert the cuvette with the
HI 93703-10
AMCO-AEPA-1
standard @10 FTU* into the
measurement cell and press the
CAL key again.