Take the HI 93703-0
bottle
containing the ZERO FTU*
Standard (or turbidity-free
dilution water) and fill the
measurement cuvette.
Note: In order to minimize any error introduced by the cuvette,
it is recommended to use, during calibration, the same cuvette
you are going to use to perform the measurement.
Insert the cuvette with the
HI 93703-0 @ ZERO FTU*
standard solution (or turbidity-
free dilution water) into the
measurement cell and press the
CAL key. A blinking “SIP
message indicates that the
instrument is performing the
measurement. After approximately
50 seconds the instrument will
ask for the HI 93703-10
standard solution @10 FTU*
by displaying “10.0“.
Repeat the same procedure with
HI 93703-10 @10 FTU*
standard solution.
Insert the cuvette with the
HI 93703-10
AMCO-AEPA-1
standard @10 FTU* into the
measurement cell and press the
CAL key again.