OPERATIONAL GUIDE
• Internal communication error*
• Internal Bus error*
* Contact the Hanna Instruments office nearest to you.
Ensure Accurate Measurements
Each time the cuvette is used, tighten the cap to the same degree.
Discard the sample soon after the reading is taken to avoid
permanently clouding the glass.
All glassware used to contain the standards and the samples
should be maintained clean, washed with the HI93703-50
cleaning solution and rinsed with HI93703-0 or turbidity-free
water.
Collect the samples in clean glass or plastic bottles, fit stoppers
and perform the analysis quickly. If necessary, store the sample in
a cool, dark place, for up to a maximum of 24 hours (the sample
needs to be kept at room temperature prior to the analysis).
To obtain a representative sample, gently, but thoroughly, mix it
before samples are taken. Do not shake (to prevent air bubbles)
and do not let the sample settle.
It is recommended to calibrate the meter with the supplied
HI93703-10 @10 FTU* standard at least once a month or more
frequently for greater accuracy.
Before inserting vials into the instrument, wipe them with
HI93703-70 or a soft, lint-free tissue. Handle vials so that no
fingerprints get on the areas where light passes (approx. the
bottom 2 cm/1 inch).
If you experience any problems in taking measurements, contact your
nearest Hanna Customer Service Center.
* 1 FTU = 1 NTU

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI93703-11 and is the answer not in the manual?

Hanna Instruments HI93703-11 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI93703-11
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals