16
BOD Parameters:
Bottle ID: a number used to identify a specific bottle.
Range: 0000 to 9999.
The type of the Sample: Sample or Seed.
Bottle Volume: the total volume of the BOD bottle.
Range: 0.1 to 300.0 mL.
Sample Volume: the volume of sample in the BOD bottle.
Range: 0.1 to 300.0 mL (for a seed sample this value is 0.0 mL and cannot be set).
Seed Volume: the volume of seed in the BOD bottle.
Range: 0.0 to 300.0 mL.
Press Prev/Next to select a different parameter on the screen.
Press / keys to modify the selected parameter’s value.
Press Save to save the BOD parameters and the initial DO, temperature, pressure and salinity values
for the specified bottle.
If a bottle with the same ID already exists, the instrument will ask for replacement confirmation.
Press Replace to replace the existing record, or ESC to return to the previous screen without replacing.
When a new record is saved the meter will display a message indicating the remaining free BOD
initial data space in %.
By pressing View ID a list of all the saved BOD initial data records will be displayed. The seed records
will have the symbol “*” displayed after the bottle ID.