33
BOD Configuration
Highlight BOD configuration.
Press Select
Parameters:
• Sample min ∆ DO ‑the minimum acceptable difference between the initial and final DO values
for a sample. If the difference is less than this value the meter will show a warning message when
evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
• Sample min end DO ‑ the minimum acceptable final DO value for a sample. If the final DO value
is less than this value the meter will show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
• Seed min ∆ DO ‑ the minimum acceptable difference between the initial and final DO values
for a seed sample. If the difference is less than this value the meter will show a warning message
when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
• Seed min end DO ‑ the minimum acceptable final DO value is less than this value the meter will
show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
SETUP
Use / keys to change the salinity value. Press Accept to confirm or ESC to exit without saving.
Press Prev/Next to select a different parameter.
Press / keys to modify the selected parameter’s value.
Press Save to save the new BOD configuration.
Press ESC to leave without changing.