EasyManua.ls Logo

Hanna Instruments HI98193 - OUR Configuration

Hanna Instruments HI98193
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
OUR Configuration
Highlight OUR configuration.
Press Select.
Parameters:
Min time ‑ the minimum time for the OUR test.
Range: 1 to 3600 seconds.
Max time ‑ the maximum time for the OUR test. The test will stop automatically when the maximum
time has elapsed.
Range: 1 to 3600 seconds.
Min start DO ‑ the minimum accepted DO value for starting the OUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
Min end DO ‑ the minimum accepted DO value at the end of the test. If the DO reading at the
end of the OUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
Total volume ‑ the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
Sample volume ‑ the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
Press Prev/Next to select a different parameter.
Press / keys to modify the selected parameter’s value.
Press Save to save the new OUR configuration.
Press ESC to leave without changing.
SETUP
www.tools.in.th
www.neonics.co.th

Related product manuals