OUR Configuration
Highlight OUR configuration.
Press Select.
Parameters:
Min time ‑ the minimum time for the OUR test.
Range: 1 to 3600 seconds.
Max time ‑ the maximum time for the OUR test. The test will stop automatically when the maximum
time has elapsed.
Range: 1 to 3600 seconds.
Min start DO ‑ the minimum accepted DO value for starting the OUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
Min end DO ‑ the minimum accepted DO value at the end of the test. If the DO reading at the
end of the OUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
Total volume ‑ the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
Sample volume ‑ the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
Press Prev/Next to select a different parameter.
Press / keys to modify the selected parameter’s value.
Press Save to save the new OUR configuration.
Press ESC to leave without changing.
SETUP
www.tools.in.th
www.neonics.co.th

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98193 and is the answer not in the manual?

Hanna Instruments HI98193 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98193
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals