EasyManuals Logo

Hanna Instruments HI98193 User Manual

Hanna Instruments HI98193
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
35
Press Select.
Min time ‑ the minimum time for the SOUR test.
Range: 1 to 3600 seconds.
Max time ‑ the maximum time for the SOUR test. The test will stop automatically when the
maximum time has elapsed.
Range: 1 to 3600 seconds.
Min start DO ‑ the minimum accepted DO value for starting the SOUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
Min end DO ‑ the minimum accepted DO value at the end of the test. If the DO reading at the
end of the SOUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
Total vol ‑ the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
Sample vol ‑ the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
Solids weight: Total solids or Volatile suspended solids weight.
Range: 0.1 to 300.0 g/L.
SOUR @ 20 °C: If this option to enabled the SOUR value is corrected to 20 °C.
Press Prev/Next to select a different parameter.
Press / keys to modify the selected parameter’s value.
Press Save to save the new SOUR configuration.
Press ESC to leave without changing.
SOUR Configuration
Highlight SOUR configuration.
SETUP
www.tools.in.th
www.neonics.co.th

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98193 and is the answer not in the manual?

Hanna Instruments HI98193 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98193
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals