Press Eval. BOD to compute the BOD for the selected bottle. If the time difference between the current
reading and the selected reading is less than 1 day the instrument will ask for record replacement
confirmation, and the BOD can’t be evaluated.
Press CFM to replace the selected records DO, temperature, pressure and salinity values with the
current values.
Press ESC to return to the previous screen without replacing.
If the conditions regarding the time difference are met, after pressing the Eval. BOD key, the instrument
will display the computed BOD value.
Press LOG to save the BOD result.
Press ESC to return to the BOD measure screen.
If the final DO reading or DO loss doesn’t meet the criteria for BOD measurements set from the
instrument’s setup, a warning message will be displayed.
Press any key to clear the warning message from the display or press HELP to view detailed infor
mation about the warning.