30
SETUP
Setup mode allows viewing and modifying the measurement parameters.
The following table lists the SETUP parameters, their valid range and the factory default settings.
Item Description Valid value Default
Calibr. Time‑out Number of days after
calibr. warning is displayed
Disabled, 1 to 7
days
Disabled
Salinity The solution’s salt content 0 to 70 g/L 0 g/L
BOD Configuration
Sample min delta
DO
Sample min end DO
Seed min delta DO
Seed min end DO
The minimum diff. between the
start and the end DO val.
The minimum end DO value
The minimum diff. between the
start and the end DO val.
The minimum end DO value
0.00 to 50.00 mg/L
0.00 to 50.00 mg/L
0.00 to 50.00 mg/L
0.00 to 50.00 mg/L
0.00 mg/L
0.00 mg/L
0.00 mg/L
0.00 mg/L
OUR Configuration
Min time
Max time
Min start DO
Min end DO
Total volume
Sample volume
The minimum time for the OUR test
The maximum time for the OUR test
The minimum DO value for starting
the OUR test
The minimum DO value at the end
of the OUR test
The total volume of the solution to
be tested
The volume of sample in the
solution to be tested
1 to 3600 sec.
1 to 3600 sec.
0.01 to 50.00 mg/L
0.00 to 50.00 mg/L
0.1 to 300.0 mL
0.1 to 300.0 mL
1 sec.
3600 sec.
0.01 mg/L
0.00 mg/L
0.1 mL
0.1 mL
SOUR configuration
Min time
Max time
Min start DO
Min end DO
The minimum time for the SOUR
test
The maximum time for the SOUR
test
The minimum DO value for starting
the SOUR test
The minimum DO value at the end
of the test
1 to 3600 sec.
1 to 3600 sec.
0.01 to 50.00 mg/L
0.00 to 50.00 mg/L
1 sec.
3600 sec.
0.01 mg/L
0.00 mg/L