50
When the reading is stable and within range of the selected calibration point, the CFM functional
key is displayed.
Press CFM to confirm.
The second expected calibration point is displayed.
Immerse the DO probe into the second vessel as close as possible to the reference thermometer.
Allow a few seconds for the probe to stabilize.
Use the / key to set the calibration point value to that measured by the reference thermometer.
When the reading is stable and within range of the selected calibration point, the CFM functional
key is displayed.
Press CFM to confirm. The instrument will return to the main screen.
TEMPERATURE CALIBRATION (for technical personnel only)