to 1mV/10mA and the output is 10mV. Assuming that the Precision is 1.5%, the output
is10.15mV maximum and 9.85mV minimum.
If the digital multimeter connected with current clamp is set to 200mV and the Precision
is 0.5%,,the readings is10.20mV maximum and 9.8mV minimum.
(10.15mV×(1+0.5%)=10.20mV,9.85mV×(1-0.5%)=9.80mV)
Accuracy:
DC Current: 1m/10mA
±(1.5%±5mA) 10mA ~ 20A
DC Current: 1mV/100mA
±(2%±20mA) 100mA ~ 40A
±(4%±0.3A) 40A ~ 65A
AC Current: 1mV/10mA
±(2%±30mA) 100mA ~ 10A (40Hz ~ 2KHz)
±(4%±30mA) 100mA ~ 10A (2KHz ~ 10KHz)
±(6%±30mA) 100mA ~ 10A (10KHz ~ 20KHz)
±(8%±30mA) 10A ~ 15A (40Hz ~ 20KHz)
AC Current: 1mV/100mA
±(2%±30mA) 100mA ~ 40A (40Hz ~ 1KHz)
±(4%±30mA) 100mA ~ 40A (1KHz ~ 2KHz)
±(6%±30mA) 100mA ~ 40A (3KHz ~ 5KHz)
±(8%±0.3A) 40A ~ 65A (40Hz ~ 20KHz)
Load Resistance: 10kΩ Typical
Temperature coefficient: 0.1×(defined accuracy) per degree C (0℃to18℃, 28℃to50℃)
SAFETY INFORMATION
The instrument complies with class II, overvoltage CAT II of the EN 61010-1, and EN
61010-2-032 standards. The pollution of Current clamp is degree 2 in accordance with
IEC 664 indoor use. If the equipment is used in a manner not specified, the protection
provided by the equipment may be impaired.
This product complies with the requirements of the following European Community
Directives: 89/336/EEC (Electromagnetic Compatibility) and 73/23/EEC (Low voltage)
as amended by 93/68/EEC (CE marking).