76 | EN
Healy GmbH | Schloss Kränzlin | Darritzer Str. 6 | 16818 Kränzlin – Germany
5. Cleaning, disinfection and maintenance
5.1 General
Maintenance activities increase the lifespan and safety of the
Healy. Cleaning and disinfection reduces the risks of cross conta-
mination and infections. Please read this section carefully.
5.2 Cleaning
Switch o the Healy before cleaning. For cleaning, commercial
cleaning products may be used. The surfaces can be wiped with
a damp sponge or cloth. Do not use too much liquid and make
sure that no liquid penetrates beneath the covers or enclosures.
Caution:
Do not use cleaning agents with solvents (e.g. acetone, nail poli-
sh, cleaners) or abrasive cleaners as they will damage the surface.
5.3 Storage between applications
After the application unplug all cables and electrodes and put
them into the storage case and store in a cool and dry location.
Refer to the safety notes for environmental storage conditions.
5.4 Maintenance
There are no serviceable parts by the users. In a case of damage,
please contact the manufacturer. The following maintenance
activities are recommended and should be considered for safe
working.
Frequency By whom Maintenance activity
Prior to each use User Check housing, cables and
electrodes for damages.
In case of damage, contact
the manufacturer or an
authorized dealer
5.5 Troubleshooting
In the following you nd typical warnings or messages that are
related to the app or hardware of the device. If you face an event
that is not listed here or a problem still exists, please contact the
manufacturer.
Event or message Cause Actions required
No connection from
the smartphone to
the device
Device is
switched o
Switch on the device
and restart the App
No connection from
the smartphone to
the device
Bluetooth in
your smart-
phone is
deactivated.
Open connectivity
settings and acti vate
Bluetooth
Repeated short
blinking of right LED
Batteries are
low
Charge the device