12. Attachment instruction for use “Electrodes” REF:
108-4004 ear electrodes (pair)
12.1 Description
Only use the ear electrodes in combination with the supplied ad-
hesive felts. In people with very dry skin, pro blems may occur due
to insucient electrical contact in individual cases. Therefore,
always ensure that the felts are soaked with physiological saline.
In the application with ear electrodes the intensity should be
set that the current is slightly perceptible. A too high current
intensity can cause dizziness and nausea. During the applica-
tion mood changes can occur, such as a feeling of heaviness or
lightness. In such cases, you should continue the application for
a few minutes until you feel normal again.
12.2 Care and storage of the electrodes
Add one to two splashes of a mild detergent to warm water.
Clean the electrode with this water and a soft cloth (or sponge)
thoroughly and rinse o after that. Do not immerse the elec-
trode in the liquid. For storage put the electrode back into the
packaging until next use and store it in a dry, clean place.
12.3 Service life of the product
The service life of the product depends on the functionality
which is in turn dependent on the number of applications, the
state of maintenance and the condition.
Upon recognition of any damages, the product shall not be used
anymore and shall be disposed.
12.4 Specications
Connectors: 2 mm connector
12.5 Important note on using the ear electrodes
Prior to the application, the electrodes must be equipped with
the supplied self-adhesive felts!
The felts must be well-moistened throughout the entire appli-
cation.
In case of any kind of discomfort, such as stinging, burning,
or pain stop the application or reduce the inten sity immedia-
tely until no unpleasant symptoms remain. If necessary, check
the contacts and improve the conductivity by remoistening
the felts..
Healy GmbH
Darritzer Str. 6
16818 Knzlin, Germany