APPENDIX
© Hermle Labortechnik GmbH Z366_V1.18_eng V
Table 1: Technical Data
Manufacturer
HERMLE Labortechnik GmbH, 78564 Wehingen
Type
Z 366
Dimensions
Width
43 cm
Depth
51 cm
Height
38 cm
Weight without rotor
52 kg
Max. Speed
15000 min
-1
Max. Volume
6 x 250 ml
Max. RCF
21379 x g
Allowable Density
1,2 kg/dm
3
Allowable Kinetic Energy
34363 Nm
Mains Power Connection AC
230 V / 50-60 Hz 1 ph
120 V / 50-60 Hz 1 ph
Voltage Fluctuation
± 10 %
± 10 %
Current Consumption
2,4 A
4,2 A
Power Consumption
530 W
520 W
Radio Interference
IEC 61326-1
Audit Requirement (BGR 500)
yes
Ambient Conditions (EN/IEC 61010-1)
-
Environment
-
High
- Ambient Temperature
- Max. Relative Humidity
-
Overvoltage Category (IEC 60364-4-443)
- Degree of Contamination
For Indoor Use Only
Use up to an altitude of 2000 m above MSL
2°C up to 35 °C
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
II
2
Class of Protection I
Class of Protection (DIN EN 60529) IP 20
I
Not suitable for use in hazardous environments.
EMV
Interference Emission , Noise
EN / IEC
61326-
1
Category B
FCC Class B
EN / IEC
61326-1
Category B
FCC Class B
Noise Level (depending on the rotor)
≤ 61 +2 dB(A)
Write from Operator
Inventory-No.:
Monitoring-No.:
Environment:
Maintenance Contract:
Responsible Service Office
HERMLE Labortechnik GmbH
or dealer service office
Siemensstra
ß
e 25
78564 Wehingen
Tel.: (49)7426 / 96 22-17
Fax: (49)7426 / 96 22-49