1 Use according to specification
The centrifuge was designed only for the separation of substances or mixtures with a density of no more than 1.2
kg/dm³, particularly of samples for preparation for human medical in-vitro diagnostics using blood collection tubes
with separation gel.
The samples centrifuged with the EBA 270 can be used directly in in-vitro diagnostic tests. Therefore, the EBA 270
centrifuge as an in-vitro diagnostic accessory is in its turn an in-vitro diagnostic medical device according to the in
vitro diagnostic medical devices regulation (EU) 2017/746.
The centrifuge is only meant for this purpose.
Another use or one which goes beyond this, is considered to be non-intended. The company Andreas Hettich GmbH
& Co. KG is not liable for damage resulting from this.
Observing all information in the operating instructions and complying with the measures described therein is also a
part of the intended use.
2 Remaining risks
The device is built according to the state-of-the-art and the recognized safety regulations. If used and handled
improperly, there could be life-threatening danger to the user or third parties, or the device could be impaired or there
could be other property damage. The device is only to be used for its intended purpose and only when it is in safe
working condition.
Malfunctions which could affect safety must be corrected immediately.
3 Technical specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model EBA 270
Basic-UDI-DI
4050674010007QC
Type 2300 2300-01
Mains voltage ( 10%) 200 - 240 V 1 100 - 127 V 1
Mains frequency 50 - 60 Hz 50 - 60 Hz
Connected load 130 VA 125 VA
Current consumption 0.7 A 1.25 A
Max. capacity 6 x 15 ml
Allowed density
1.2 kg/dm
3
Speed (RPM) 4000
Force (RCF) 2254
Kinetic energy 250 Nm
Obligatory inspection
(DGUV Regel 100 - 500)
no
Ambient conditions (EN / IEC 61010-1)
 Set-up site
Indoors only
 Altitude
Up to 2000 m above sea level
 Ambient temperature
2°C to 40°C
 Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
 Excess-voltage category
(IEC 60364-4-443)

 Pollution degree
2
Device protection class
Not suitable for use in explosion-endangered areas.
EMC
 Emitted interference,
Interference immunity
EN / IEC 61326-1,
Class B
FCC Class B
Noise level (dependent on rotor)
51 dB(A)
Dimensions
 Width
326 mm
 Depth
389 mm
 Height
239 mm
Weight approx. 13.5 kg