EN
32/45 Rev. 01 / 11.2022 AB1101EN_SA
20.8 Querying the hours of operation and the number of centrifugation runs
The hours of operation are divided into internal and external hours of operation.
Internal hours of operation: Total time the device was switched on.
External hours of operation: Total time of the previous centrifugation runs.
With the rotor at a standstill, the query can proceed as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key.
Press and hold the button for eight
seconds.
PROG t/min:sRPM
Press the key until the following is
displayed:
PROG t/min:sRPM
START
PULSE
Press the key.
TimeExt : External hours of operation.
Example:
PROG t/min:sRPM
Press the key.
TimeInt : Internal hours of operation.
Example:
PROG t/min:sRPM
Press the key.
Starts : Number of all centrifugation runs.
Example:
PROG t/min:sRPM
STOP
OPEN
Press the key twice to exit the "
Time &
Cycles" menu or press three times to exit
the "
MACHINE MENU ".
PROG t/min:sRPM
Example:
PROG t/min:sRPM