  • J
    Jessica StevensSep 23, 2025
    Why does my Hettich ROTOFIX 32 A Laboratory Equipment display a Version Error?
    • K
      Kevin ReedSep 23, 2025
      A Version Error on your Hettich Laboratory Equipment means that the wrong machine version is set, causing the control section to skip to the Settings menu. To correct this, use the [t] buttons to set the number 7, press the [STOP] button to save the settings, and then perform a MAINS RESET.
  • O
    onguyenAug 25, 2025
    What to do if Hettich ROTOFIX 32 A Laboratory Equipment displays error '-2-'?
    • R
      richardellisAug 25, 2025
      If your Hettich Laboratory Equipment shows error '-2-', it means there was a loss of mains power during the centrifugation run, and the run was not completed. You should open the lid and press the [START] button. If necessary, repeat the centrifugation run.
  • M
    Mackenzie JohnsonAug 25, 2025
    How to fix no display on Hettich Laboratory Equipment?
    • C
      Christine ShawAug 26, 2025
      If your Hettich Laboratory Equipment's display is not working, it could be due to no power or defective mains input fuses. Check the supply voltage and the mains input fuse. Also, ensure the mains switch is set to [I].
  • A
    Alexander ShahAug 27, 2025
    What does error '-4-' mean on Hettich Laboratory Equipment and how to fix it?
    • M
      Michael GarciaAug 27, 2025
      If your Hettich Laboratory Equipment displays error '-4-', it signals a communication error in either the control or power section. Perform a MAINS RESET to resolve this issue.
  • A
    allen28Aug 27, 2025
    What to do if Hettich ROTOFIX 32 A Laboratory Equipment displays error '-5-'?
    • D
      Donald OdomAug 27, 2025
      If your Hettich Laboratory Equipment shows error '-5-', it indicates an overload, possibly due to a defective motor or motor control. Try performing a MAINS RESET to address this issue.
  • R
    Ronald McleanSep 16, 2025
    What does error -1- mean on my Hettich Laboratory Equipment?
    • C
      Christina MendezSep 16, 2025
      Error -1- on your Hettich Laboratory Equipment indicates a tacho error, caused by speed pulse failure. Make sure that the device is not switched off while the ‘Rotation’ indicator is lit up and rotating. Wait until the ‘Lid closed’ symbol is displayed (after approximately 100 seconds) and then perform a MAINS RESET.
  • J
    John Austin MDSep 13, 2025
    What does error -F- mean on my Hettich Laboratory Equipment?
    • A
      amy94Sep 13, 2025
      Error -F- on your Hettich Laboratory Equipment indicates no rotor detection when starting, meaning no rotor is inserted or the tacho is defective. To attempt to fix this, perform a MAINS RESET.
  • S
    stephaniecampbellSep 15, 2025
    Why is there no speed indicator on my Hettich Laboratory Equipment?
    • D
      Donna GomezSep 15, 2025
      The absence of a speed indicator on your Hettich Laboratory Equipment may be due to a version error, where the wrong machine version is set, causing the control section to skip to the Settings menu. To fix this, use the [t] buttons to set the number 7, then press the [STOP] button to save the settings, and finally perform a MAINS RESET.
  • W
    William DavisSep 5, 2025
    What does error -4- mean on my Hettich ROTOFIX 32 A?
    • T
      Tiffany WilsonSep 5, 2025
      Error -4- on your Hettich Laboratory Equipment indicates a communication issue within the control or power section. To solve this, perform a MAINS RESET.
  • A
    Alisha AlexanderSep 3, 2025
    What does error -6- mean on my Hettich ROTOFIX 32 A?
    • A
      arthur71Sep 3, 2025
      Error -6- on your Hettich Laboratory Equipment signifies overvoltage, meaning the mains voltage is outside of the acceptable range. Try performing a MAINS RESET and check the mains voltage.

Summary

Safety

Intended Use of Centrifuge

Defines the centrifuge's intended purpose and application.

User Personnel Requirements

Details user qualifications and requirements.

General Safety Instructions

Provides critical safety warnings and procedures.

Centrifuge Commissioning

Setup and Connection

Guide to setting up and connecting the centrifuge.

Centrifuge Operation

Powering On and Off

Procedure for powering the centrifuge on/off.

Rotor Installation and Removal

Guide for safe rotor removal and installation.

Loading the Centrifuge

Guidelines for loading centrifuge tubes into rotors/buckets.

Centrifugation Procedures

Procedures for performing centrifugation runs.

Software Operation

Setting Centrifugation Parameters

How to set and adjust centrifugation parameters.

Cleaning and Care

Cleaning and Disinfection

Detailed instructions for cleaning and disinfecting the device.

Routine Maintenance

Procedures for routine maintenance.

Troubleshooting Guide

Fault Description and Remedies

Table of common faults, causes, and remedies.

Performing a MAINS RESET

Steps to perform a main power reset.