AB1206en_SARev.: 06 / 11.2023
34 / 39
Troubleshooting
Replacing centrifuge tubes
The maximum permissible number of run cycles or the expiry date can
be seen marked on the accessories.
CAUTION
Risk of injury from broken glass.
Broken glass may cause glass splinters and contaminated
liquids to be found inside the centrifuge.
Wear cut-resistant gloves.
Wear protective goggles and a face mask.
Broken parts of the tube, glass splinters and spilled centrifuge material
must be removed completely in the event of leakage or if a centrifuge
tube breaks. Glass splinters that are not removed will cause further glass
breakage.
The rubber inserts and the plastic sleeves of the rotors must be replaced
after a glass breakage.
Disinfection must be carried out if the material is infectious.
9
Troubleshooting
9.1
Fault description
Customer service must be notied if the fault cannot be rectied based on
the fault table. State the centrifuge type and serial number. Both numbers
can be seen on the type plate of the centrifuge.
Fault description Cause Remedy
no display No power. Mains input fuses
defective.
Check the supply voltage.
Check the mains input fuse.
Set the mains switch to
[I]
.
-1- Tacho error. Speed pulse
failure.
The device must not be switched off
while the
‘Rotation’
indicator is lit up
and rotating.
Wait until the
‘Lid closed’
symbol is
displayed (after approx. 100 seconds)
and then perform a MAINS RESET.
-2- Loss of mains power during
the centrifugation run. The
centrifugation run was not
completed.
Open the lid and press the
[START]
button.
If required: Repeat the centrifugation
run.
-3- Imbalance. The rotor is
unevenly loaded.
Open the lid.
Check the loading of the rotor.
Repeat the centrifugation run.
-4- Communication. Error in the
control section or power sec-
tion.
Perform a MAINS RESET.
-5- Overload. Motor or motor con-
trol defective.
Perform a MAINS RESET.