Fig. 21: Mains input fuse
1 Fuse holder
2 Snap lock
The mains fuses are located next to the mains switch.
The mains switch is in switch position
[O]
1. Disconnect the mains cable from the device plug.
2. Press the snap lock (
2
) against the fuse holder (
1
) and pull it out.
3. Replace the defective mains input fuses.
Only use fuses with the nominal value specied for the type: see the
table below.
4. Push in the fuse holder (
1
) until the snap lock engages.
5. Reconnect the device to the mains.
Model Type Fuse Order no.
ROTOFIX 32 A 1206, 1206-34 T 3.15 AH/250
V
E997
ROTOFIX 32 A 1206-01,
1206-33
T 5 AH/250 V E914
10
Disposal
10.1
General instructions
The device can be disposed of via the manufacturer.
A Return Material Authorisation (RMA) form must always be
requested for a return.
If necessary, contact the Technical Service Department of
the manufacturer.
−
Andreas Hettich GmbH & Co.KG
−
Föhrenstrasse 12
− 78532 Tuttlingen, Germany
− Phone: +49 7461 705 1400
− E-mail: service@hettichlab.com
WARNING
Risk of pollution and contamination for people and the envi-
ronment.
When disposing of the centrifuge, people and the environ-
ment may be polluted or contaminated by incorrect or
improper disposal.
− Removal and disposal may be carried out only by a
trained and authorised service personnel.
The device is intended for the commercial sector ("Business to Business" -
B2B).
According to Directive 2012/19/EU, the devices may no longer be disposed
of with household waste.
The devices are assigned to the following groups according to the Stiftung
Elektro-Altgeräte Register (EAR (German foundation under civil law)):
■ Group 5 (small devices)