Troubleshooting
Fault description Cause Remedy
-6- Overvoltage. Mains voltage
outside tolerances.
Perform a MAINS RESET.
Check the mains voltage.
-7- Overspeed. Error in the power
section.
Perform a MAINS RESET.
-8- Undervoltage. Mains voltage
outside tolerances.
Perform a MAINS RESET.
Check the mains voltage.
-9- Overtemperature. Overtemper-
ature switch in the motor has
tripped.
Open the lid using the emergency
release.
Let the motor cool down.
Version Error Wrong machine version set.
The control section skips to
the Settings menu.
Use the
[t]
buttons to set the number
7.
Press the
[STOP]
button to save the
settings.
Perform a MAINS RESET.
no speed indicator. Machine
version set in the time indi-
cator.
Version Error. Wrong machine
version set. The control sec-
tion skips to the Settings
menu.
Use the
[t]
buttons to set the number
7.
Press the
[STOP]
button to save the
settings.
Perform a MAINS RESET.
-c- Controller watchdog. Error in
the power section.
Perform a MAINS RESET.
-d- Lid lock error.
Perform a MAINS RESET.
-E- Short circuit in the control sec-
tion / power section.
Perform a MAINS RESET.
-F- No rotor detection when
starting. No rotor inserted or
defective tacho.
Perform a MAINS RESET.
rot... A new rotor has been
detected.
Press the
[START]
button.
All indicator seg-
ments light up.
-
Notify customer service.
9.2
Perform a MAINS RESET
1. Set the mains switch to
[0]
.
2. Wait 10 seconds.
3. Set the mains switch to
[I]
.
9.3 Emergency release
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.

Table of Contents