AB1206en_SARev.: 06 / 11.2023
30 / 39
Cleaning and care
7.3.1.2 Setting an audible signal
1.
Switch off the mains switch.
2.
Press and hold down the
[RPM/RCF x 100]
button and the
[IMPULSE]
button simultaneously.
3. Switch on the mains switch and release the buttons.
4. Press the
[RPM/RCF x 100]
button repeatedly until
‘BEL 1’
or
‘BEL
0’
is displayed.
5. Use the
[t]
buttons beneath the time indicator to set
‘0’
or
‘1 ’
.
0 = audible signal disabled.
1 = audible signal enabled.
6. Press the
[STOP]
button to save the settings.
8 Cleaning and care
8.1 Overview table
Chap. Task to execute
if required
daily
weekly
Annually
Page
8 Cleaning and care 30
8.3 Cleaning 31
8.3 Cleaning the device X 31
8.3 Cleaning the biosafety systems X 31
8.3 Cleaning the accessories X 32
8.4 Disinfection 32
8.4 Disinfecting the device X 32
8.4 Disinfecting the accessories X 32
8.5 Maintenance 33
8.5 Greasing the rubber seal of the centrifuging
chamber
X 33
8.5 Greasing the rubber seal of the biosafety system X 33
8.5 Trunnion greasing X 33
8.5 Checking the accessories X 33
8.5 Checking the biosafety system X 33
8.5 Inspecting the centrifuging chamber for damage X 33
8.5 Greasing the motor shaft X 33
8.5 Accessories with a limited service life X 33